Maintenance
Confirming Interlock Circuit
Perform the following procedure to confirm that the interlock circuit works
correctly.
1. Connect an interlock connection cable between the
Intlk
terminal and the
interlock terminal of an instrument such as the Keysight 4155/4156 and E5270.
2. Close the lid of the 16442B.
3. Set all channel output to 0 V.
4. Confirm that the High Voltage LED is turning off.
5. Change a channel output to greater than 42 V. The High Voltage LED has to turn
on.
6. Open the lid of the 16442B. The High Voltage LED has to turn off.
Opening the lid will make the instrument cause an error or change voltage to
0 V.
Servicing
This section provides information for trained service personnel to repair the test
fixture.
provides the Keysight part numbers of replaceable parts. When
a part needs to be replaced, order the parts from the nearest Keysight Technologies
Sales and Service Office.
WARNING
High voltages may be present in the test fixture when voltage or current is
added to the test fixture. Dangerous voltage may be generated on wires or parts
and the danger of electric shock exists.