2
Specifications
16
Keysight V281C/D and W281C/D Operating and Service Manual
Environmental Requirements
Temperature — What To Watch Out For
Due to the critical dimensions and tight tolerances of the 1.0 mm connectors on your adapters,
electrical characteristics will change with temperature. The operating temperature is a critical
factor in the performance during measurements and between calibrations.
Table 2-1
Environmental requirements
Parameter
Required values/ranges
Operating temperature
0 °C to 55 °C (32 °F to 131 °F)
Storage temperature
–40 °C to 75 °C (–40 °F to 167 °F)
Altitude
– Operation
– Storage
–
|
15,000 meters (
_
50,000 feet)
–
|
15,000 meters (
_
50,000 feet)
Relative humidity
– Operation
– Storage
Always non-condensing
– 0 to 80% (26 °C maximum dry bulb)
– 0 to 90%
NOTE
Remember
your fingers are a heat source, so avoid handling the devices unnecessarily during
calibration.