Using the U3050BH11 with N5245A
Using the U3050BH11 with N5245A
This section describes how to operate the Test Set with the PNA-X. The Test Set does not
require any control signals. Power levels are controlled by the PNA-X.
The Test Set will amplify and combine Source Out Port 1 and Port 3 signals and route the signal
to the U3050BH11 Port 1. The Port 1 Reference Mixer Switch must be set to
“External” for the RCVR R1 IN to function.
The front panel bulkhead access connectors allow for custom configurations and minimizes
connections to the PNA-X. Refer to
The power level from Port 1 on the PNA-X must not exceed −10 dBm. The
power level from Port 3 on the PNA-X must not exceed 0 dBm due to the gain
of the amplifier. This is to ensure that the PNA-X RCVR A IN power level is not
exceeded (+15 dBm).
These levels can be exceeded if attenuation is placed
before the PNA-X RCVR A IN port.
Figure
11 Bypass Switch Select
Port 1 Reference
Mixer Switch
14
Keysight U3050BH11 User's and Service Guide
Содержание U3050BH11
Страница 4: ...Contents 2...
Страница 5: ...U3050BH11 Keysight U3050BH11 User s and Service Guide 1...
Страница 21: ...U3050BH11 Operational Check Figure 13 Keysight U3050BH11 User s and Service Guide 17...
Страница 23: ...U3050BH11 Operational Check Figure 15 Keysight U3050BH11 User s and Service Guide 19...
Страница 25: ...U3050BH11 Operational Check Figure 17 Keysight U3050BH11 User s and Service Guide 21...
Страница 27: ...U3050BH11 Operational Check Figure 19 Keysight U3050BH11 User s and Service Guide 23...
Страница 29: ...U3050BH11 Operational Check Figure 21 Keysight U3050BH11 User s and Service Guide 25...
Страница 31: ...U3050BH11 Operational Check Figure 23 Keysight U3050BH11 User s and Service Guide 27...
Страница 33: ...U3050BH11 Operational Check Figure 25 Keysight U3050BH11 User s and Service Guide 29...
Страница 37: ...U3050BH11 Theory of Operation Figure 27 U3050BH11 Block Diagram Keysight U3050BH11 User s and Service Guide 33...