U3047AM08
Controlling the Test Set
-
RF Path Configuration
Figure 19
RF Path Configuration
If the analyzer has Option 029 or H29 (noise figure capability) verify that
the path configuration is set to Default. If the network analyzer is not in
Default configuration, ports 5 and 9 will not operate correctly. Select
Trace/Chan > Channel > Hardware Setup > Path Config...
-OR-
(Instrument > Setup > Internal Hardware > RF Path Config...)
and in the
drop-down menu select
Default > OK
. Save this configuration as a User
Preset by selecting
Save > User Preset > Save current state as User
Preset
. Do not use the factory Preset (User Preset Off),
o
r
the analyzer
will return to Option 029 path configuration.
Keysight U3047AM08 User's and Service Guide
29