External Voltage Supply Configuration
Figure 41 below illustrates an example of the connection with an external DC power supply.
Connect the positive and negative voltage supply from the external power supply to the
positive input (pin 13) and the negative input (pin 14). Connect the power supply DC ground to
pin 15.
Turning On the Test Set using an External Power Supply
1. Turn On the Test Set.
2. Connect the DUT.
3. Turn On the external power supply.
Turning Off the Test Set using an External Power Supply
1. Turn Off the Power Supply.
2. Turn Off the Test Set.
3. Disconnect the DUT.
Figure
41
Test Set to the DUT and External DC Power Supply
8
1
15
14
13
TO DUT (Only required Lines)
GND
+
Gnd
Gnd
DC Power Supply
12
13
14
15
1
2
8
2 to 5 Vdc
TEST SET
Line 1
Line 2
Line 8
GND
Positive: 0 to 5 Vdc
Negative: 5 to 0 Vdc
NC
9
(Rear Panel View)
Keysight U3024AH10 User's and Service Guide
52
Содержание U3024AH10
Страница 5: ...U3024AH10...
Страница 31: ...Figure 17 24 Port RF Interface Cable Configuration Keysight U3024AH10 User s and Service Guide 26...
Страница 52: ...Figure 38 Example 1 Test Set Signal Path 47 Keysight U3024AH10 User s and Service Guide...
Страница 81: ...System Block Diagrams Figure 56 14 Port System Configuration Keysight U3024AH10 User s and Service Guide 76...
Страница 82: ...Figure 57 24 Port System Configuration 77 Keysight U3024AH10 User s and Service Guide...
Страница 83: ...Figure 58 2 Way High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 78...
Страница 84: ...Figure 59 Odd Port High Power Configuration 1 Watt 79 Keysight U3024AH10 User s and Service Guide...
Страница 85: ...Figure 60 Port 1 High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 80...
Страница 95: ...Servicing Diagram Figure 67 Test Set Electrical Diagram Keysight U3024AH10 User s and Service Guide 90...
Страница 105: ......