Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines using address and data
values. Refer to
for more detailed information.
The <address> <data> commands will only function while the PNA is in
Standalone Mode. Select
Restart as a standalone PNA
>
OK
. Refer to
All DUT control lines are initially configured in a High logic state after the line power has been
turned on, and until set by the <address>.<data> input. The control lines will remain in the last
state set when changing from Multiport mode to Standalone mode. Sending a command 112.0
will ensure that all of the control lines are high.
As described in
“Setting the Test Port Paths with Address and Data Inputs,”
<address>.<data> value are determined in a similar manner, with the exception that Table 14
(below) will be used. The <address> value is always = 112
Example: To change the default DUT control line, so that lines 1 & 8 are changed to the
alternate position, enter front panel PNA-X control mode line entry = 112.129 > OK. All other
lines remain in the default High state.
Another method that can be used to change the logic state of these control lines is
explained in
“External Test Set Control Operation” on page 32
Only the top Test Set control lines operate in a 24-Port system.
Table 1
4
Test Set DUT Control Address and Data Logic Table
<address>
<data>
Description
Line/
Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic low or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8
Keysight U3024AH10 User's and Service Guide
48
Содержание U3024AH10
Страница 5: ...U3024AH10...
Страница 31: ...Figure 17 24 Port RF Interface Cable Configuration Keysight U3024AH10 User s and Service Guide 26...
Страница 52: ...Figure 38 Example 1 Test Set Signal Path 47 Keysight U3024AH10 User s and Service Guide...
Страница 81: ...System Block Diagrams Figure 56 14 Port System Configuration Keysight U3024AH10 User s and Service Guide 76...
Страница 82: ...Figure 57 24 Port System Configuration 77 Keysight U3024AH10 User s and Service Guide...
Страница 83: ...Figure 58 2 Way High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 78...
Страница 84: ...Figure 59 Odd Port High Power Configuration 1 Watt 79 Keysight U3024AH10 User s and Service Guide...
Страница 85: ...Figure 60 Port 1 High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 80...
Страница 95: ...Servicing Diagram Figure 67 Test Set Electrical Diagram Keysight U3024AH10 User s and Service Guide 90...
Страница 105: ......