1
Calibration Procedures
Performance Verification Tests
28
U1450 Series/U1461A Service Guide
3
DCV
Turn the rotary
switch to the
position.
6 V
6 V
±0.006 V
±0.006 V
60 V
60 V
±0.06 V
±0.06 V
600 V
600 V
±0.6 V
±0.6 V
1000 V
1000 V
±2 V
±2 V
4
DCmV
Turn the rotary
switch to the
position.
60 mV
60 mV
-
±0.064 mV
600 mV
600 mV
-
±0.64 mV
[1] The accuracy is specified after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
5
ACV
Turn the rotary
switch to the
position.
6 V
6 V, 65 Hz
±0.063 V
±0.063 V
6 V, 5 kHz
±0.093 V
±0.093 V
6 V, 20 kHz
±0.124 V
±0.124 V
60 V
60 V, 65 Hz
±0.63 V
±0.63 V
60 V, 5 kHz
±0.93 V
±0.93 V
60 V, 20 kHz
±1.24 V
±1.24 V
600 V
600 V, 65 Hz
±6.3 V
±6.3 V
600 V, 999 Hz
±9.3 V
±9.3 V
1000 V
1000 V, 65 Hz
±13 V
±13 V
1000 V, 999 Hz
±18 V
±18 V
6
ACV (LPF)
While the
rotary switch is in the
position, press
once.
6 V
6 V, 439 Hz
-
±0.363 V
60 V
60 V, 439 Hz
-
±3.63 V
600 V
600 V, 439 Hz
-
±36.3 V
1000 V
1000 V, 439 Hz
-
±63 V
Table 1-3
Performance verification tests for models U1461A/U1453A (continued)
Step
Test function
Range
5520A/5522A
output
Error from nominal 1 year
U1453A
U1461A
Содержание U1450 Series
Страница 1: ...Keysight U1450 Series Insulation Testers and U1461A Insulation Multimeter Service Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X U1450 Series U1461A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...XIV U1450 Series U1461A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...XVI U1450 Series U1461A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...