Instrument Information
Power requirements
The only physical installation of your Keysight signal analyzer is a connection
to a power source. Line voltage does
not
need to be selected.
This analyzer does
not
contain customer serviceable fuses.
AC power cord
The analyzer is equipped with a three-wire power cord, in accordance with
international safety standards. This cable grounds the analyzer cabinet when
connected to an appropriate power line outlet. The cable appropriate to the
original shipping location is included with the analyzer.
The instruments can operate with mains supply voltage fluctuations up to
± 10% of the nominal voltage.
This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor inside or outside of the product is
likely to make the product dangerous. Intentional interruption is
prohibited. (IEC 348 clauses 17.3.3c & 17.3.4)
Failure to ground the analyzer properly can result in personal injury.
Before turning on the analyzer, you must connect its protective earth
terminals to the protective conductor of the main power cable. Insert the
main power cable plug into a socket outlet that has a protective earth
contact only. DO NOT defeat the earth-grounding protection by using an
extension cable, power cable, or autotransformer without a protective
ground conductor.
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 61010 Second Edition and IEC 664 respectively.
This instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
The Mains wiring and connectors shall be compatible with the connector
used in the premise electrical system. Failure, to ensure adequate earth
grounding by not using the correct components may cause product
damage, and serious injury.
Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord can
cause product damage.