Installation Note N9030-90077
31
Option BBA, Analog Baseband I/Q Inputs Upgrade Kit
Verify the Installation
1.
Turn on the analyzer and wait for the instrument to fully boot up.
2.
Verify the spectrum analyzer application loads and sweeps as expected. Press
System
,
Show
Hard ware
on the analyzer and verify that the Main board and the
identify
themselves as:
—
BBIQ
—
BBIQ Interface
3.
Press
System
,
Show System
on the analyzer to display a list of all installed options. You
should see N9030B-BBA.
Before performing adjustments and performance verification tests using the N7814A Calibration
and Adjustment Software for Keysight RF and Microwave Instruments, press
MODE/MEAS
,
I/Q
Analyzer (Basic)
,
OK
. Press
Input/Output
, Select
Input
,
I/Q
,
I/Q
,
I/Q Cable Calibrate...
on the
analyzer and follow the on-screen directions.
If the newly installed hardware is functioning properly the cable cal should work fine. Press
MODE/MEAS
,
Spectrum Analyzer
,
Swept SA
,
OK
. Press
Input/Output
,
Select Input
,
RF
.