26
Installation Note N9030-90098
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Remove Outer Case
1.
Disconnect the instrument from ac power.
2.
Refer to
Figure 13
. Using the T-20 driver, remove the four screws (two on each side)
(1)
that
attach the handle strap
(2)
on each side of the instrument.
3.
Using the T-20 driver, remove the four screws (including washers)
(3)
that hold the rear feet
(4)
in place.
4.
Pull the instrument cover
(5)
off towards the rear of the instrument.
Figure 13
Standard Instrument Outer Case Removal