background image

14

Installation Note 

N9000-90032

Option T03, T06 - Tracking Generator Installation Kit

6. Installation Verification

Power the instrument on and verify that the instrument will not only boot up, but that the signal 
analyzer application software will load properly. This would include verifying that there are no error 
messages displayed in the Status Bar of the application window.

Adjustments and Performance Verification Tests

Adjustments and performance verification testing requires the use of the calibration software. The 
latest software information and downloads are available at:

http://www.agilent.com/find/calibrationsoftware

Adjustments Required  

TG Level Accuracy

TG Path Delay

Performance Testing Recommended  

TG Absolute Amplitude and Flatness

TG Power Sweep

TG Spurious Outputs

End of Installation

  

For assistance, contact your nearest Agilent Technologies Sales and Service Office. To find your local 
Agilent office access the following URL, or if in the United States, call the following telephone 
number:

http://www.agilent.com/find/assist

1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)

Содержание N9000-90032

Страница 1: ...Notice This document contains references to Agilent Please note that Agilent s Test and Measurement business has become Keysight Technologies For more information go to www keysight com Installation Note Part Number N9000 90032 Printed in USA July 2014 N9000 90032 N9000 90032 ...

Страница 2: ...to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose H VLJKt Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Copyright 2011 201 H VLJKt Technologies Inc ...

Страница 3: ...is document will guide you through the following steps which will need to be followed closely and performed in the order defined 1 Pre upgrade Status Check 2 Hardware Installation 3 Update the Instrument Firmware 4 Install the License Files 5 Function Activation 6 Installation Verification Products Affected CXA N9000A Serial Numbers CXA To Be Performed By X Keysight Service Center X Personnel Qual...

Страница 4: ...erformance Tests and Adjustment Software Microsoft Windows based personal computer with internet access and USB port USB storage device 2GB free memory Quantity Description Agilent Part Number 1 Tracking Generator Assembly N9000 65402 1 RF Input Connector Assembly N9000 60032 1 EMI O ring Shielding 8160 1637 1 Cable Assembly SMA TG RF 2nd LO N9000 20104 1 Cable Assembly SMA TG RF Bridge N9000 2010...

Страница 5: ...oceeding with this upgrade or make detailed notes of the instrument status in the space provided below and address them once this upgrade has been completed if they continue to be present ______________________________________________________________________________ ____________________________________________________________________________ ________________________________________________________...

Страница 6: ...er case of the instrument a Refer to Figure 1 Using the T 20 driver remove the four screws two on each side 1 that attach the handle strap 2 on each side of the instrument b Using the T 20 driver remove the four screws including washers 3 that hold the rear feet 4 in place c Pull the instrument cover 5 off towards the rear of the instrument Figure 1 Standard Instrument Outer Case Removal 3 Remove ...

Страница 7: ...b To Remove the reference bracket 2 use the T 10 driver to remove the six screws 5 including two at the rear panel attaching the reference bracket to the instrument and the four screws 4 attaching the top brace to the reference bracket Figure 2 Top Brace and Reference Bracket Removal ...

Страница 8: ...or Installation Kit 4 Refer to Figure 3 to remove the front frame assembly Figure 3 Front Frame Assembly Removal a Remove the overlay 2 on the front frame assembly b Remove the eight screws 1 four on each side to detach the front frame assembly from the chassis 1 2 ...

Страница 9: ...00 90032 9 Option T03 T06 Tracking Generator Installation Kit c Pull the front frame assembly carefully away from the chassis Refer to Figure 4 to remove the ribbon cable W1 from the mother board Figure 4 Front Panel Cable ...

Страница 10: ... the chassis and press down to plug it into the motherboard Figure 5 Tracking Generator Assembly Install the RF Input Connector Assembly 1 Locate the tracking generator assembly in the retrofit kit 2 Refer to Figure 6 to install the RF input connector assembly 1 into the instrument chassis 3 Tighten the screws 3 with 9 inch pounds 4 Install the EMI O ring 2 Shielding onto the RF input connector Fi...

Страница 11: ... to RF input connector assembly and tighten the SMA connector with 10 in lbs Figure 7 Install the RF Semi rigid Cable 3 Refer to Figure 8 connect the other end of the cable 1 to tracking generator assembly Tighten the the SMA connector with 10 in lbs Figure 8 Install the Cable Assembly Attach the semi rigid cable to RF connector assembly tighten the SMA connector with 10 in lbs Semi rigid cable N9...

Страница 12: ...gure 8 install N9000 20105 2 N9000 20106 3 N9000 20104 4 and N9000 20107 5 onto the corresponding instrument assemblies 6 Re install the front frame assembly top brace reference bracket and instrument outer case Avoid pinching cables when installing the top brace 7 Locate the overlay in the retrofit kit Refer to Figure 9 attach the overlay 2 to the front frame assembly 1 Figure 9 Cable Assembly 1 ...

Страница 13: ...e and email with an attached License File 2 Located the USB Storage Device Perform a virus scan on this device before use Save the License File to the root directory of the USB Storage Device 3 Connect the USB Storage Device to the signal analyzer USB port 4 The signal analyzer will automatically consume the License File This may take a few minutes When the License File is consumed the Agilent Lic...

Страница 14: ...erformance verification testing requires the use of the calibration software The latest software information and downloads are available at http www agilent com find calibrationsoftware Adjustments Required TG Level Accuracy TG Path Delay Performance Testing Recommended TG Absolute Amplitude and Flatness TG Power Sweep TG Spurious Outputs End of Installation For assistance contact your nearest Agi...

Отзывы: