Adjustments
3
Keysight N8262A Service Guide
61
Power Reference Level Adjustment
Equipment
As per the test equipment list for the
1 mW Power Reference Level Test
Test method
1
Set: Coarse = 834, Fine = 550
2
Measure the power ref. level as per the performance test:
a
If the result is >1 mW, then increment COARSE by 1.
b
If the result is <1 mW, then decrement COARSE by 1.
3
Repeat
until the result crosses the 1 mW boundary (in either direction).
4
Measure power ref. level as per the performance test:
a
If the result is >1 mW, then decrement FINE by 1.
b
If the result is <1 mW, then increment FINE by 1.
5
Repeat
until the result crosses the 1 mW boundary (in either direction).
6
The adjustment is completed.
NOTE
– This adjustment can be performed manually via the commands:
SERV:CAL:ADJ:COUR
<Value>
SERV:CAL:ADJ:COUR?
SERV:CAL:ADJ:FINE
<Value>
SERV:CAL:ADJ:FINE?
(Refer to the
Programming Guide
for further details on the use of these
commands)
– COARSE and FINE values are valid in the range of 0 to 1023
– If adjustment is not possible, then a fault may be present in the DUT (see
Chapter 5, "Troubleshooting Guide"
).
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Service Guide ...
Страница 18: ...18 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 40: ...1 Specifications 40 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 58: ...2 Performance Tests 58 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 62: ...3 Adjustments 62 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...4 Theory of Operation 70 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...