Keysight N778xC Series Polarization Instruments Programming Guide
59
Measurement Operations & Settings
4
Command:
:TRIGger:CONFiguration?
Syntax:
:TRIGger:CONFiguration?
Description:
Returns the hardware trigger configuration.
Parameters:
None
Parameters:
0 or DISabled:
1 or DEFault:
2 or PASSthrough:
3 or LOOPback:
4 or SCRambler2polarimeter:
5 or POLarimeter2scrambler:
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming trigger response for each
slot.
A trigger at the Input Trigger Connector generates a trigger at the Output Trigger
Connector automatically. No triggers from the instrument at the output.
The same as PASSthrough. This is included for compatibility reasons.
The internal polarization scrambler forward the trigger-out signal to the built-in
polarimeter. This signal can be delayed by configure the
:PCONtroller:SEQuence:HOLDoff command).
The internal polarimeter forward the trigger-out signal to the built-in scrambler.
Example:
:trig:conf?
->
DEF
Affects
All modules (SCR and POL are only available at the N7786C and N7788C which have a build-in polarization scrambler and
polarimeter).
Command:
:TRIGger:CONFiguration
Syntax:
:TRIGger:CONFiguration
Description:
Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.
Parameters:
0 or DISabled:
1 or DEFault:
2 or PASSthrough:
3 or LOOPback:
4 or SCRambler2polarimeter:
5 or POLarimeter2scrambler:
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming trigger response for each
slot.
A trigger at the Input Trigger Connector generates a trigger at the Output Trigger
Connector automatically. No triggers from the instrument at the output.
The same as PASSthrough. This is included for compatibility reasons.
The internal polarization scrambler forward the trigger-out signal to the built-in
polarimeter. This signal can be delayed by configure the
:PCONtroller:SEQuence:HOLDoff command).
The internal polarimeter forward the trigger-out signal to the built-in scrambler.
Response:
none
Example:
:trig:conf dis
Affects
All modules (SCR and POL are only available at the N7786C and N7788C which have a build-in polarization scrambler and
polarimeter).