86
N7020A and N7024A Power Rail Probes User’s Guide
8
Performance Verification
f
Set the channel offset for the oscilloscope to
-640mV
and the trigger level to
-640mV
as displayed in
Figure 32
.
g
The f2 trace now has the frequency response with the N7024A probe
attached. The already saved and displayed memory 1 trace has the
frequency response of the coax cable. For comparison, grab the f2 trace and
align it with the memory 1 trace, if needed as displayed in
Figure 32
.
h
The frequency response of the N7014A probe is the difference between the
f2 trace and memory 1 trace. Verify whether or not this difference is less
than
3db
at 6 GHz or below. If it is less, then the probe passes the test.
Notice that in the figure below, the N7024A probe passed the bandwidth
verification test as the frequency response of the probe s not above 3db at 6
GHz or below.
Figure 32
Frequency response of the N7024A probe
Содержание N7020A
Страница 1: ...Keysight N7020A N7024A Power Rail Probes User s Guide ...
Страница 18: ...18 N7020A and N7024A Power Rail Probes User s Guide 1 N7020A N7024A Probes Overview ...
Страница 34: ...34 N7020A and N7024A Power Rail Probes User s Guide 3 Specifications and Characteristics ...
Страница 38: ...38 N7020A and N7024A Power Rail Probes User s Guide 4 Safety and Regulatory Information ...
Страница 54: ...54 N7020A and N7024A Power Rail Probes User s Guide 5 Using the Power Rail Probe Figure 11 ESD Workstation ...
Страница 60: ...60 N7020A and N7024A Power Rail Probes User s Guide 6 Calibrating ...
Страница 76: ...76 N7020A and N7024A Power Rail Probes User s Guide 7 Performance Plots ...
Страница 90: ...90 N7020A and N7024A Power Rail Probes User s Guide 8 Performance Verification ...
Страница 92: ...92 N7020A and N7024A Power Rail Probes User s Guide Index ...