Single Unit Connections
Additional Load Considerations
AC Mains Considerations
Keysight APS models have a bi-directional three-phase AC input converter, which allows for bi-
directional power flow between the AC mains and DC output terminals. Power flow is 100 % into the
unit when sourcing power, and up to 10% into the AC mains when sinking power. In a standard power
supply, energy only flows from the AC to the DC output terminals. In a regenerative power supply,
energy also flows from the DC output terminals back to the AC mains whenever the unit is sinking
current, either from rapid down-programming of the output or from discharging an energy source such
as a battery. This return of energy back to the AC mains is referred to as regenerative operation.
The AC input converter of the APS unit employs firmware algorithms to maintain high power factor and
low total harmonic distortion across a wide range of operating conditions. For more information refer
to the AC input portion of the
section.
Follow all wiring instructions for the power cord provided in this section of the manual. For example,
the APS unit employs a connection to AC neutral. The neutral current is very small, but the instrument
will not turn on if this connection is not provided..
Exceeding the AC input voltage rating or mis-wiring the line connection can create
permanent damage in the following devices.
The following components are used to provide immunity against line-to-line transients as high as 2 kV,
and line-to-ground transients as high as 4 kV:
l
300 Vrms voltage limiting devices are installed between the line and neutral connections.
l
600 Vrms voltage limiting devices are installed between the line-to-line connections.
l
A gas-discharge tube is installed between the neutral connection and chassis ground.
Several internal protection functions are used to safeguard the instrument against abnormal line
conditions while also preventing AC mains islanding. The following conditions cause the instrument to
shut down:
Keysight N69100 Series Operating and Service Guide
57
2 Installing the Instrument
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......