N5511A-K26 Rack Mount Instructions
Rack Mount Ventilation
When installing the instrument(s) into a cabinet consideration shall be given to the convection flow
into and out of the cabinet. Consideration shall also be given to the individual instruments to avoid
having the heated discharge of one instrument, now becoming the cooling intake air for another
instrument.
Another area of concern is verification that the maximum ambient operating temperature of the
instrument(s) is not exceeded by cabinet installation. Keysight recommends forced air convection
whenever an instrument(s) are installed in a cabinet and further recommends that the maximum
operating temperature of the cabinet be reduced 10°C from the lowest, of the maximum operating
temperature of a single instrument.
If there are any concerns or special requirements a Keysight Field Engineer should be consulted to
assure instrument(s) temperature compliance and performance.