Service Information
2-
Measure LO OUT (Gain and Power)
Prepare the PNA to measure S12 with a normalization calibration. Refer to
“PNA Normalization Calibration” on page
. Use a frequency range of 8 to
20 GHz and a power level of
−
10 dBm. The cable on PNA Port 2 must be
long enough to reach the rear panel of the test set.
1.
Connect the PNA Port 2 cable to the rear panel LO IN.
2.
Connect PNA Port 1 to Port 1 LO OUT on the test set. Verify the LO Output
gain is
> 20 dB. The trace should be within 10 dB of “flat” and have no
significant
holes or discontinuities. Refer to
3.
Select [Sweep] > Sweep Type > CW Time. Set the sweep time to
10
seconds.
4.
5.
Disconnect the cable from LO IN and measure the power at the end of the
cable using a power sensor. The power should be in the
−
8 to
−
12 dBm
range. Adjust the PNA power output level if needed.
Reconnect the cable
to LO IN.
6.
Measure the Port 1 LO OUT power using a power meter. The power should
be +12 dBm
(± 1.5 dB). Measure LO OUT for the other ports on the test set.
Figure 2-
1
4
LO Output Gain with
−
10 dB Input
Move PNA Port 2 cable to Port 2 LO Out and confirm a similar result
.
7
2
Keysight N5261/62A User’s & Service Guide
Содержание N5261A
Страница 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Страница 6: ......
Страница 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Страница 51: ...Keysight N5261 62A User s Service Guide 45 Installation System Operation Figure 1 30 Uncalibrated S12 and S21...
Страница 54: ...48 Keysight N5261 62A User s Service Guide...
Страница 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...
Страница 63: ...Service Information 2 Figure 2 8 Sloped Attenuator Cable Assembly Keysight N5261 62A User s Service Guide 57...