Receiver Power Check
Power accuracy is dependent the accuracy of your power sensor.
After calibration, the reference receivers will show power at the test ports.
Power Level Control and Max Power Level Check
1. Set up two traces, making sure "Statistics" is turned on.
a. For 1-Port and 2-Port: a1,1 and a2,2 (or R1,1 and R2,2 receiver)
b. For 3-Port and 4-port: a3,3 and a4,4 (or R3,3 and R4,4 receiver)
2. Measure the power level at +2 dBm (default). Verify the traces are measuring 2 dBm
(mean), ± 1 dB.
3. Set the power level to 0 dBm. Verify the traces are measuring 0 dBm (mean), ± 1 dB.
4. Set the power level to 13 dBm. Verify the traces are measuring a mean of 13 dBm,±1
dBm.
5. See
for a display example for 2-port, or
for 4 -port.
N5252-90002 User's Guide