1
Introduction
10
N432A User’s Guide
Rear Panel Connections
No.
Connection
1
Ground connector
2
Recorder output terminal 1
Recorder output connection is made via a BNC connector. This output produces a DC
voltage that corresponds to the power level of the channel input.
3
AC inlet
The N432A has an autoconfiguring power supply. This allows it to operate over a range of
voltages without manually being set to a certain voltage.
4
V
RF
and V
COMP
output terminals
The V
RF
BNC terminal outputs the RF bridge voltage, while the V
COMP
BNC terminal outputs
the compensation bridge voltage. Both the V
RF
and
V
COMP
outputs are used for calibrating
the N432A and for precision power measurements.
5
USB Mini-B port
Allows the N432A to be controlled remotely over the USB interface
6
LAN interface
Allows the N432A to be controlled remotely over the LAN interface
7
GPIB interface
Allows the N432A to be controlled remotely over the General Purpose Interface Bus
(IEEE-488) standard interface
5
6
4
7
2
1
3
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......