![Keysight Technologies N432A Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/keysight-technologies/n432a/n432a_user-manual_1962025054.webp)
2
Operating Information
32
N432A User’s Guide
If you have modified the resistance type and the Ruser value, the
subsequent power- up setting will correspond to the
state
setting.
Procedure
Use the following steps to set the resistance type.
1
Press
to access the channel setup screen.
2
Use the up/down arrow key to highlight the
Res Type
value field, and
press
to display the Res Type pop- up list. Use the up/down
arrow key to highlight the desired bridge resistance type, and press
to enter the selection.
Figure 2-3
Resistance type pop-up list
3
When Ruser is selected, you can enter a resistance value within a
±
10%
tolerance level of the current bridge resistance setting. Use the left/right
arrow key to highlight the resistance value field, then press
to
display the Ruser pop- up. Use the numeric keypad to enter the
required value and press
to confirm the entry.
Figure 2-4
Ruser pop-up
Enter
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......