3
Maintenance
104
N432A User’s Guide
To read the error queue from the front panel:
•
press
>
and use
to scroll through each error
message.
To read the error queue from the remote interface, use:
•
the
SYSTem:ERRor?
query.
Error queue messages have the following format:
Error queue message
For example, –113,"Undefined header;*T<Err> ST?".
Errors are retrieved in a first- in, first- out (FIFO) order. If more than 30
errors occur, the error queue overflows and the last error in the queue is
replaced with error –350,“Queue overflow”. Any time the queue overflows,
the most recent errors are discarded.
When the errors are read, they are removed from the error queue. This
opens a position at the end of the queue for a new error message, if one
is subsequently detected. When all errors have been read from the queue,
further error queries 0,“No error”.
To delete all the errors in the queue from the front panel:
•
press
>
and use
.
To delete all the errors in the queue remotely, use:
•
the
*CLS
(clear status) command.
The error queue is also cleared when the N432A power has been switched
off.
Error List
Next
Error
”
;
Device-
Error
Description
Dependent
Info
Number
“
,
Error List
Clear Errors
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......