15.

Set

 Calibration Factor Switch (dB) on PM5B front panel to 

0.3 dB 

for WR6.5

 and

WR8.0 (see table in 

VDI PM5B Operational Manual - Appendix 2

 for current

 

listed

values). Note: Connect WR10 flanges of sensor and taper together for best
match.

 The smaller opening should face your DUT.

16.

Connect sensor to DUT, set frequency to 110 GHz, and power level to 0 dBm
(1.00 mW), RF OFF.

17.

Ensure DUT power is off. On the front panel of the PM5B, press "Zero."

18.

Turn RF on and measure power. Measure 1.000 mW ± your accuracy in setting
the power level and ± 0.050 mW accuracy.

PM5B calorimeter amplitude response is flat at ± 5% across a frequency band

 

of 75 - 110 GHz.

Calibrated at DC to NIST

 standards

.

Only the loss of the added tapers requires a Cal Factor dB setting, otherwise it

 

remains at 0.

2

 dB

 for W-Band measurements

.

A "Zero" of the PM5B should be done after the calorimeter is at a stable

 

temperature.

Measurement should be recorded after the reading stabilizes

 (see response

 

time below). 

Important 

Notes:

Scale

90% Response Time

1

Table 

2

 

Typical PM5B Performance

200 mW

 

20 mW

 

2 mW

 

200 µW

0.15 seconds

0.2 seconds

0.6 seconds

12 seconds

RMS Noise

~ 0.02 µW

~ 0.08 µW

0.03 µW

0.003 µW

1. Response time is given as the time from the application of an input to a response at the analog output of

90% of the final reading.

Keysight N1913PM5B Product Note

9

Содержание N1913PM5B

Страница 1: ...Product Note Keysight Technologies N1913PM5B Calorimetric Power Meter ...

Страница 2: ...ent customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the Software The EULA and the license set forth therein does not require or permit among o...

Страница 3: ...icture are optional accessories and are not provided unless ordered with the PM5B The power line cord shown is ordered for the country of use and is supplied outside the Pelican case as an accessory Keysight N1913PM5B Product Note 3 ...

Страница 4: ... and up to 1 5 THz with taper accessory USB connectivity Code compatible with N522xB PNA or N524xB PNAX with firmware A 13 80 08 Includes Lab VIEW 7 express operating software that can be loaded into the PNA PNAX or computer Includes built in calibration verification feature PM5B Operational Manual https vadiodes com images Products PowerMeter PM5manual VDI 7241_PM5B_Manual pdf VDI Erickson Power ...

Страница 5: ... 1 dB RF accuracy typical 5 Frequency 75 110 GHz up to 1 5 THz with a taper A table in the VDI PM5B Operational Manual Appendix 2 provides the values for the Cal Factor setting if an additional taper is added The optional waveguide tapers have a flat frequency response so only one Cal Factor value is set on the PM5B front panel for the waveguide band The loss of the WR10 sensor and taper values ar...

Страница 6: ...ower supply and if using software the USB cable to the rear panel on the PM5B power meter See Figure 1 and Figure 2 below Figure 1 N1913PM5B with Sensor and Cable Connected to the Front Panel Figure 2 N1913PM5B with Power Supply and USB Cable Connected to the Rear Panel The USB connection is usually very snug and may require some additional pressure to insert or remove completely 6 Keysight N1913P...

Страница 7: ...to re zero display Display Reads power on selected range Sensor Head Connector Port For use with standard Amphenol cable 4 Turn on the PM5B power meter using the power ON OFF switch on the rear panel Figure 4 5 Start the PM5B control program optional not required with manual operation 6 Enter serial number for example 520V optional not required with manual operation Figure 4 PM5B Power Meter Rear ...

Страница 8: ...n on RF and measure the power set on the source Measure 1 000 mW your accuracy in settling the power level and 0 050 mW estimated accuracy Measurement Example WR10 Sensor with pre installed WR10 Section 110 GHz 1 mW 0 dBm 14 Connect a WR6 5 or WR8 0 taper to the WR10 sensor Do not remove the WR10 section Measurement Example WR6 5 or WR8 0 Taper 110 GHz 1 mW 0 dBm Figure 5 Sensor Head Waveguide and...

Страница 9: ...rimeter amplitude response is flat at 5 across a frequency band of 75 110 GHz Calibrated at DC to NIST standards Only the loss of the added tapers requires a Cal Factor dB setting otherwise it remains at 0 2 dB for W Band measurements A Zero of the PM5B should be done after the calorimeter is at a stable temperature Measurement should be recorded after the reading stabilizes see response time belo...

Страница 10: ... not covered by the product warranty Nominal nom A general descriptive term that does not imply a level of performance Description Table 3 PM5B Specifications Summary Specification Maximum VSWR typ Frequency Response nom Input Power Range typ Maximum Input Power spec Cal Factor Adjustment nom Calibration Check nom Temperature Drift Compensation typ Data Port nom Display nom Analog Output nom AC Po...

Страница 11: ... Mechanical Dimensions Size 5 1 x 4 8 x 7 6 cm 8 9 x 15 2 x 26 7 cm Sensor Power meter Specification Table 5 PM5B Environmental Specifications Range 10 30 C Operating temperature spec Keysight N1913PM5B Product Note 11 ...

Страница 12: ...NA or VNA is Figure 6 USB Cable Connection to the Rear Panel of the Power Meter Figure 7 USB Cable Connection to the Rear Panel of the Analyzer The USB connection is usually very snug and may require some additional pressure to insert or remove completely A 13 80 08 or above 1 Turn on the power meter then connect the USB cable from the rear port of the power meter to the rear panel of the analyzer...

Страница 13: ...ront panel press the CAL key to bring up the screen shown in Figure 8 Figure 8 Press the CAL key 3 On the display screen press the Other Cals softkey Figure 9 Figure 9 Select Other Cals Keysight N1913PM5B Product Note 13 ...

Страница 14: ...n select the Source Power Cal softkey Figure 10 Figure 10 Select Source Power Cal 5 On the display screen select the Power Meter Config button Figure 11 Figure 11 Select Power Meter Config 14 Keysight N1913PM5B Product Note ...

Страница 15: ...tings window comes up select Any and in the VISA resource string alias bar type ASRL3 INSTR as shown in Figure 12 Figure 12 Power Meter Settings 7 Select OK to begin the power meter calibration Keysight N1913PM5B Product Note 15 ...

Страница 16: ...16 Keysight N1913PM5B Product Note ...

Страница 17: ...this information the Keysight representative can determine whether your unit is still within its warranty period To verify the contents shipped with your product refer to the Box Content List included with the shipment Inspect the shipping container If the container or packing material is damaged it should be kept until the contents of the shipment have been checked mechanically and electrically I...

Страница 18: ...This information is subject to change without notice Keysight Technologies 2020 Edition 2 July 2020 Supersedes March 2020 www keysight com ...

Отзывы: