background image

Find us at www.keysight.com 

 

 

 

 

 

 

 

 

 

Page 7 

Step 2-2. Select the cable if you utilized an interlock circuit 

The M9602A/M9603A is equipped with an interlock safety feature to protect against exposure to 
voltages above a user-specified value. You can engage the safety lock using interlock pins on the 
connector located at the front panel. Normally, these pins are routed to a shielding box or test fixture 
that must be closed to complete the interlock circuit. 

You can install an interlock circuit as shown in Figure 3. If your shielding box has a BNC connector for 
the interlock circuit, you can use the PX0101A-001 or 002 BNC to ferrule terminal cable with a 
connector-terminal block to connect the M96

02A/M9603A’s interlock pins to it. For more detailed 

information, please refer to the Keysight M9602A/M9603A Startup Guide. 

 

Step 2-3. Select the cables if you use an external trigger 

There are trigger lines in the PXIe chassis available to synchronize the M9602A/M9603A with the other 
PXIe modules. In addition, there are two external trigger lines available if you need to synchronize it with 
an external instrument. You can use the PX0101A-001 or 002 BNC to ferrule terminal cable with a 
connector-terminal bl

ock to connect the M9602A/M9603A’s external trigger lines to the external 

instrument. 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 5. An example of interlock circuit installation 

 

PX0101A-001 or 002 BNC 
to ferrule terminal cable 

Содержание M9602A

Страница 1: ...ptions and associated accessories This guide provides step by step instructions to help you configure the product to meet the desired test requirements Model number M9602A PXIe SMU 15 MSa s 1 pA 60 V...

Страница 2: ...hort bar 1ea Short bar to connect low terminals to chassis common orderable part number is M9601 87001 includes 5 pcs Connector terminal block 2 5 mm 6 terminal 1ea Connector terminal block to connect...

Страница 3: ...1 High current triaxial cable2 Triaxial to SMB cable3 50 s or DC High current triaxial cable High current triaxial cable Triaxial to SMB cable Two wire High current triaxial cable NA NA 1 A low induct...

Страница 4: ...connection is required for low current measurement 10 nA or less measurement Do not connect the outer shield or Low Force to the chassis frame ground of the DUT when M9603A is grounded Figure 2 Connec...

Страница 5: ...NC f adaptor instead PX0105A 001 002 Low inductance BNC cable 1 5 m 3 m 1 High force N1254A 106 Triaxial m to BNC f adaptor 1 Used with PX0105A Four wire connection for pulse 50 s or DC Figure 3 shows...

Страница 6: ...e 4 shows the connection for 2 wire connection For this configuration select the items shown in Table 3 Table 3 Required items for 2 wire connection Product number Description Qty Note PX0104A 001 002...

Страница 7: ...X0101A 001 or 002 BNC to ferrule terminal cable with a connector terminal block to connect the M9602A M9603A s interlock pins to it For more detailed information please refer to the Keysight M9602A M9...

Страница 8: ...tion of calibration comes as standard The optional ISO 17025 not accredited ANSI Z540 and commercial calibration certificate with test data are available as follows Description Product number Addition...

Страница 9: ...Connection between M9602A M9603A and your DUT Two wire connection or four wire connection If you are measuring very small resistances or applying a very large current you should use the four wire mea...

Страница 10: ...urements 1 nA requires a guarding to prevent leakage through the measurement cable Figure 7 illustrates a simplified overview of the guarding technique Guarded measurements require the use of triaxial...