background image

Find us at www.keysight.com 

 

 

 

 

 

 

 

 

 

Page 6 

Table 2. Required items for 4-wire connection for pulse (> 50 µs) or DC 

Product number 

Description 

Qty. 

Note 

PX0103A-001/002 

Triaxial to SMB cable 
1.5 m/ 3 m 

Low sense 

PX0104A-001/002 

High current triaxial cable, 4 A 
1.5 m/ 3 m 

High force, High sense 

 

Two-wire connection 

Figure 4 shows the connection for 2-wire connection. For this configuration, select the items shown in 
Table 3.  

 

Table 3. Required items for 2-wire connection 

Product number 

Description 

Qty. 

Note 

PX0104A-001/002 

High current triaxial cable, 4 A 
1.5 m/ 3 m 

High Force 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 4. Connection for 2-wire connection. 

S

 

DUT Interface 

DUT 

In the case of  
Low force = floating 

If you perform grounded measurement, 
attach the short bar. 
 
Or if you perform floating measurement, 
remove the short bar. 

PX0104A 
High current triaxial cable 

Содержание M9602A

Страница 1: ...ptions and associated accessories This guide provides step by step instructions to help you configure the product to meet the desired test requirements Model number M9602A PXIe SMU 15 MSa s 1 pA 60 V...

Страница 2: ...hort bar 1ea Short bar to connect low terminals to chassis common orderable part number is M9601 87001 includes 5 pcs Connector terminal block 2 5 mm 6 terminal 1ea Connector terminal block to connect...

Страница 3: ...1 High current triaxial cable2 Triaxial to SMB cable3 50 s or DC High current triaxial cable High current triaxial cable Triaxial to SMB cable Two wire High current triaxial cable NA NA 1 A low induct...

Страница 4: ...connection is required for low current measurement 10 nA or less measurement Do not connect the outer shield or Low Force to the chassis frame ground of the DUT when M9603A is grounded Figure 2 Connec...

Страница 5: ...NC f adaptor instead PX0105A 001 002 Low inductance BNC cable 1 5 m 3 m 1 High force N1254A 106 Triaxial m to BNC f adaptor 1 Used with PX0105A Four wire connection for pulse 50 s or DC Figure 3 shows...

Страница 6: ...e 4 shows the connection for 2 wire connection For this configuration select the items shown in Table 3 Table 3 Required items for 2 wire connection Product number Description Qty Note PX0104A 001 002...

Страница 7: ...X0101A 001 or 002 BNC to ferrule terminal cable with a connector terminal block to connect the M9602A M9603A s interlock pins to it For more detailed information please refer to the Keysight M9602A M9...

Страница 8: ...tion of calibration comes as standard The optional ISO 17025 not accredited ANSI Z540 and commercial calibration certificate with test data are available as follows Description Product number Addition...

Страница 9: ...Connection between M9602A M9603A and your DUT Two wire connection or four wire connection If you are measuring very small resistances or applying a very large current you should use the four wire mea...

Страница 10: ...urements 1 nA requires a guarding to prevent leakage through the measurement cable Figure 7 illustrates a simplified overview of the guarding technique Guarded measurements require the use of triaxial...