Find us at www.keysight.com
Page 6
Table 2. Required items for 4-wire connection for pulse (> 50 µs) or DC
Product number
Description
Qty.
Note
PX0103A-001/002
Triaxial to SMB cable
1.5 m/ 3 m
1
Low sense
PX0104A-001/002
High current triaxial cable, 4 A
1.5 m/ 3 m
2
High force, High sense
Two-wire connection
Figure 4 shows the connection for 2-wire connection. For this configuration, select the items shown in
Table 3.
Table 3. Required items for 2-wire connection
Product number
Description
Qty.
Note
PX0104A-001/002
High current triaxial cable, 4 A
1.5 m/ 3 m
1
High Force
Figure 4. Connection for 2-wire connection.
S
DUT Interface
DUT
In the case of
Low force = floating
If you perform grounded measurement,
attach the short bar.
Or if you perform floating measurement,
remove the short bar.
PX0104A
High current triaxial cable