136
Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
4
Remote Programming
:INSTrument:FPGA:MODE
Command
:INSTrument:FPGA:MODE {NORMal|OPENfpga}
Description
The command selects the DPU FPGA mode.
NORMal
The default DPU FPGA image contained in the software
installation is used.
OPENfpga
The default DPU FPGA image or a user-provided image can be
used.
The query returns the DPU FPGA mode.
Returned Format
[:INSTrument:FPGA:MODE] <fpga_mode><NL>
:INSTrument:FPGA:UPDate
Command
:INSTrument:FPGA:UPDate {DPUA|DPUB}, <filename>
Description
The command loads a new image into the DPU FPGA.
DPUA
Selects DPU A for update.
DPUB
Selects DPU B for update.
<file_name>
File name of the FPGA update image
:INSTrument:FPGA:RESet
Command
:INSTrument:FPGA:RESet {DPUA|DPUB}
Description
The command loads the default image into the DPU FPGA.
DPUA
Selects DPU A for update.
DPUB
Selects DPU B for update.