6000L Series Oscilloscopes Service Guide
2
Testing Performance
Overview
To completely test and troubleshoot the mixed-signal oscilloscope, you will create
and use a test connector accessory, as described in this chapter.
• The test connector is only required for oscilloscopes that have the MSO option
licensed (enabled).
• The connector is used in the digital channel threshold accuracy test.
• The test connector makes it easy for you to connect the oscilloscope probes to
function generators and measurement equipment with minimum electrical
distortion.
Let the Equipment Warm Up Before Testing
For accurate test results, let the test equipment and the oscilloscope warm up 30
minutes before testing.
Verifying Test Results
During the tests, record the readings in the Performance Test Record on
.
To verify whether a test passes, verify that the reading is within the limits in the
Performance Test Record.
Conventions
The following conventions will be used when referring to oscilloscope models
throughout this chapter.
If a performance test fails
If a performance test fails, first perform the User Cal procedure
Table 1
Conventions
Models
Referred to as:
DSO6014L
100 MHz Models
DSO6054L
500 MHz Models
DSO6104L
1 GHz Models
Содержание InfiniVision 6000L Series
Страница 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Страница 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Страница 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Страница 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Страница 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...