614
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
24
:POWer Commands
:POWer:CLResponse:WGEN:VOLTage:PROFile
Command Syntax
:POWer:CLResponse:WGEN:VOLTage:PROFile {{0 | OFF} | {1 | ON}}
The :POWer:CLResponse:WGEN:VOLTage:PROFile command enables or disables
the ability to set initial waveform generator ramp amplitudes for each frequency
range.
With amplitude profiling, you can use lower amplitudes at frequencies where the
device under test (DUT) is sensitive to distortion and use higher amplitudes where
the DUT is less sensitive to distortion. Power supply feedback networks are
typically most sensitive near the 0 dB cross-over frequency.
You can often observe distortions during the test. If the input test sine wave begins
to look lopsided, clipped, or somewhat triangular in shape (nonsinusoidal), you are
probably encountering distortion due to overdriving your DUT. Optimizing test
amplitudes to achieve the best dynamic range measurements is often an iterative
process of running your frequency response measurements multiple times.
Query Syntax
:POWer:CLResponse:WGEN:VOLTage:PROFile?
The :POWer:CLResponse:WGEN:VOLTage:PROFile? query returns the voltage
profile setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also
•
•
•
•
":POWer:CLResponse:FREQuency:MODE"
•
":POWer:CLResponse:FREQuency:STARt"
•
":POWer:CLResponse:FREQuency:STOP"
•
•
":POWer:CLResponse:SOURce:INPut"
•
":POWer:CLResponse:SOURce:OUTPut"
•
•
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...