40
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
To verify bandwidth (-3 dB)
This test checks the bandwidth (-3 dB) of the oscilloscope. In this test you will use
a signal generator and a power meter.
Table 13
Bandwidth (-3 dB) Test Limits
Models
Test Limits
1 GHz Models
All channels (-3 dB), dc to 1 GHz
500 MHz Models
All channels (-3 dB), dc to 500 MHz
350 MHz Models
All channels (-3 dB), dc to 350 MHz
200 MHz Models
All channels (-3 dB), dc to 200 MHz
100 MHz Models
All channels (-3 dB), dc to 100 MHz
70 MHz Models
All channels (-3 dB), dc to 70 MHz
Table 14
Equipment Required to Verify Bandwidth (-3 dB)
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
100 kHz - 1 GHz at 200 mVrms
Keysight N5171B
Power Meter
1 MHz - 1 GHz ±3% accuracy
Keysight N1914A
Power Sensor
1 MHz - 1 GHz ±3% accuracy
Keysight E9304A or N8482A
Power Splitter
outputs differ by < 0.15 dB
Keysight 11667A
Cable
Type N (m) 24 inch
Keysight 11500B
Adapter
Type N (m) to BNC (m)
Keysight 1250-0082 or
Pomona 3288 with Pomona 3533
50 Ohm Feedthrough
Termination
50
Ω
BNC (f) to BNC (m), when testing 2000 X-Series
oscilloscopes
Keysight 0960-0301
Содержание InfiniiVision 2000
Страница 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Страница 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Страница 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...