Troubleshooting
4
90000 A-Series Service Guide
67
Front Panel Display Troubleshooting
Connect external monitor to VGA port. Turn unit on.
Does
display appear
on ext monitor?
No
Yes
Is
front panel
black?
Check display cable
connection to display card
and LCD
No
Does
front panel
display
work?
Try a golden display
card to verify failure
Does
display work
with golden
card?
Replace LCD
Replace display
card
Replace
motherboard
Check inverter board control
cable
Yes
Check voltage on pin 1 of
J3 on display card. Should
be 12V.
Is
voltage
OK?
Replace LCD
Yes
Does
front panel
display
work?
Replace
inverter
No
Front Panel Display Trouble Isolation
Go to “Verifying Basic Operation”
Yes
No
Yes
No
No
Yes
Replace
display card
Содержание Infiniium 90000A Series
Страница 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Страница 6: ...6 90000 A Series Oscilloscopes Service Guide...
Страница 12: ...12 90000 A Series Service Guide 1 General Information...
Страница 18: ...18 90000 A Series Service Guide 2 Calibration...
Страница 110: ...110 90000 A Series Service Guide 5 Replacing Assemblies P owerharness between backplaneand m otherboard O n O ff ca ble...
Страница 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Страница 119: ...Replaceable Parts 6 90000 A Series Service Guide 119 Exploded Views Front Frame and Front Panel...
Страница 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Страница 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Страница 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Страница 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Страница 135: ......