Installation Note N9010-90041
Option EXM, External Mixing Upgrade Kit
Tools Required
— T-10 TORX Driver
— T-20 TORX Driver
— 5/16-inch torque wrench
— ¼-inch open-end wrench
— Diagonal cutters
— Keysight Calibration and Adjustment Software, N7814A (revision E.10.00 or later)
— Test equipment and computer supported by the X- Series Performance Tests and Adjustment
Software
— EXA Signal Analyzer Service Guide. This manual is available as:
— N9010A Option OBW or
— Keysight part number N9010-90050
— Microsoft Windows based personal computer with internet access and USB port
— USB storage device with > 2 GB free memory
Initial Instrument Functionality Check
Power on the instrument and allow the instrument to boot up. Run an alignment and display the
measurement screen. (The instrument will probably display a spectrum analyzer screen and you will
see the instrument sweeping.)
There should be no alignment failures. If there are failures, investigate and fix the problem before
continuing.
WARNING
Before you disassemble the instrument, turn the power switch to
Standby. After the instrument has completely shut down, unplug the
instrument. Failure to unplug the instrument can result in personal
injury.
CAUTION
Electrostatic discharge (ESD) can damage or destroy electronic components.
All work on electronic assemblies should be performed at a static-safe
workstation. Refer to the documentation that pertains to your instrument for
information about static-safe workstations and ordering static-safe
accessories.
Installation Procedure
1.
Connect a power cord to the analyzer and turn on the analyzer.