i3070 Lean Inline System (E9988EL)
Replacing the Power Supply
11
4
Open the F-3 fuse holder and replace the 2.5 A slow blow fuse with a new 6.0 A
slow blow fuse.
5
Remove old power supply:
If installing the Omron 20 A AC/DC power supply, remove PS1 and PS2.
If installing the Cosel 6.4 A AC/DC power supply, remove PS3.
a
Open the protective cover and loosen the screws to release all cables at the
AC/DC power supply terminal point of the power supply to be replaced.
b
Remove the button screws and any cable ties securing the power supply to
be replaced. Remove the power supply and bracket.
F-3
protective cover
PS1
PS3
PS2