Preparation for Accurate Measurement (Executing Correction)
Executing Offset Correction
4. Pr
ep
ar
at
ion
fo
r
Accu
ra
te
M
eas
ur
em
en
t
(Executing
C
orrection)
Executing Offset Correction
Turning ON/OFF the correction function
When you turn ON the offset correction, assuming that the measured value before
correction is
and the offset correction data is
, the measured value is
compensated as
.
To turn ON/OFF the offset correction, use the following command.
•
[:SENSe]:CORRection:OFFSet[:STATe] on page 243
You cannot turn ON/OFF the primary parameter and secondary parameter separately.
However, if you set the correction value to 0, the state is actually the same as OFF even if
the offset correction is ON. Therefore, in practice you can realize separate ON/OFF states
by setting the correction value for either parameter to 0.
Setting up the correction data
To set up the offset correction data, use the following command.
•
[:SENSe]:CORRection:OFFSet:DATA on page 242
The entered value is set as the offset correction data for the measurement frequency at the
time of the entry.
Meas
Offset
Meas Offset
–
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...