Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:OFFSet:DATA
Syntax
[:SENSe]:CORRection:OFFSet:DATA <numeric 1>,<numeric 2>
[:SENSe]:CORRection:OFFSet:DATA?
Description
Sets the correction values for the primary parameter and secondary parameter used in the
offset correction function.
These correction data are measured for the measurement frequency used in executing the
command (set with the
command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
NOTE
The unit of parameter changes depending on the limit range designation method.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
[:SENSe]:CORRection:OFFSet[:STATe] on page 243
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - OFFSET - A|B
<numeric 1>
<numeric 2>
Description
The offset correction value for the
primary parameter.
The offset correction value for the
secondary parameter.
Range
-999.999 to 999.999
-99.9999E9 to 99.9999E9
Initial value 0
0
Unit
F (farad)
Depends on the setup of the
secondary parameter.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...