216
Chapter 10
Command Reference
E4981A SCPI Commands
:DISPlay[:WINDow]:TEXT2[:DATA]:FMSD[:STATe]
Syntax
:DISPlay[:WINDow]:TEXT2[:DATA]:FMSD[:STATe] {ON|OFF|1|0}
:DISPlay[:WINDow]:TEXT2[:DATA]:FMSD[:STATe]?
Description
Determines whether to use the fixed point display when displaying both the primary and
secondary parameters. In the following cases, the fixed point display is always used
regardless of this setting (the value of the highest digit is also fixed).
• When displaying deviation as a percentage in the deviation measurement mode.
• When displaying the results of D, Q and % of the secondary parameter.
Parameters
Query response
{1|0}<newline><^END>
Related commands
:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD:DATA on page 213
:CALCulate1:MATH:EXPRession:NAME on page 189
:CALCulate1:MATH:STATe on page 190
:CALCulate2:MATH:EXPRession:NAME on page 192
:CALCulate2:MATH:STATe on page 193
Equivalent key
sequence
[
Display Format
] - <Secondary Measurement Parameter i.e D/Q/G/Rp/Rs> - D.P.
AUTO|D.P. FIX
Description
ON or 1
Specifies the fixed point display.
OFF or 0 (initial value)
Specifies the floating point display.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...