Troubleshooting 5
Series E4360 Service Guide
57
Power Module Troubleshooting Sheet 2
Calibration
fixed problem?
Voltage/current
accuracy off?
Cannot get to
full power?
Remote sensing
problems?
1. Replace DPM
2. Replace main board
1. Check output connector
2. Replace main board
No
No
Yes
No
Yes
Yes
No
End
Spec failure?
Calibrate unit
From Sheet 1
Yes
No
Yes