90
Keysight CX3300 User’s Guide, Edition 5
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A
Using the CX1206A Sensor Head
The CX3300 cannot acquire the user calibration data of the CX1101A connected
to the CX1206A. Initially, the Gain value is set to 1 and the Offset value is set to 0.
If you want to use the user calibration capability, you need to obtain the Gain and
Offset values and enter the values to the CX3300 manually.
To obtain the user calibration data, you need a current source which supports
10
A output.
This section describes how to obtain the user calibration data by using the
following equipment and accessories.
•
Keysight B2900 (B2901A/B2902A/B2911A/B2912A) or equivalent
•
BNC-to-banana adapter, 2 ea.
•
BNC cable, 24 inches, 1ea., furnished with the CX3300
Figure 3-2
Connection Diagram
CX1101A
CX1206A's
high current adapter
CX1206A's
expander
CX1206A's
H
xtension cable
BNC-to-banana adapter
BNC cable
Ground marker
BNC-to-banana adapter
Current source
CX3300