Keysight CX3300 Series Device Current Waveform Analyzer
User’s Guide
3
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A 88
User calibration allows you to perform error correction of measurement data. This
is effective for making more accurate measurements. This chapter describes how
to perform the user calibration.
The user calibration must be performed after warming-up of 30 minutes.
The user calibration is effective for:
•
the environmental temperature of 0 to 40
C
•
3
C from the temperature when the last user calibration is performed
•
24 hours since the last user calibration is performed
All of the user calibration data is cleared by applying the factory default. The data
is not cleared by turning the CX3300 off.