background image

07 | Keysight | Low Current Semiconductor Measurements

 

Using the B2980A Series Ammeter - Technical Overview

Example 1. Monitoring LED Forward diode current 

(continued)

The following steps outline how to setup and measure the low current I/V characteristics of an LED.

Detailed B2985A front panel setup instructions

Step 1. Press the lower-right [View] key to show the View menus in the function keys below the screen. Next, press the [Meter View]  

 

function key, and then press the [AMPS(I)] assist key at the right of the screen.

 

Note: In case of the B2981A/83A the instrument boots-up into this state and these operations are not necessary.

Step 2. Check if the Voltage Source ield is set to 0 V. If not then follow the steps shown below to set the output voltage to zero volts. 

Step 3.  (1) Press the Voltage Source [On/Off] key to enable the voltage source to output 0 V.

 

Note: If using the B2981A/83A then set the external voltage source to zero volts.

 

(2) Press the Ammeter [On/Off] key to connect the ammeter.

 

(3) A single current measurement is made automatically.

(1) Press the              key to show the View function key.        

(2) Press                key, then 

(3) press              key.

1.

2.

3.

(3) The field pointer changes to green (EDIT mode), 
and the status information will change to EDIT

.

(4) Set the voltage to zero volts by rotating the          ,

and press          to fix the value.

Status

2.

1.

3.

(1) If the Voltage Source value is not 0 V, then 
perform the following steps.

(2) Rotate           and  to move the field pointer onto 
the Voltage Source  value, and press the knob to 
edit the voltage. 

(1) Voltage Source

On

(2) Ammeter On

(3) The first measurement is made automatically.

1.

2.

Содержание B2981A

Страница 1: ...ght Technologies Low Current Semiconductor Measurements Using the B2980A Series Ammeter Keysight B2981A B2983A Femto Picoammeter Keysight B2985A B2987A Electrometer High Resistance Meter Technical Ove...

Страница 2: ...lexibility 3 The ability to improve data quality through averaging iltering and zero cancelation 4 Integrated capability to analyze measurement data via histograms time domain graphs etc The Keysight...

Страница 3: ...rovide stable measurements and a lexible coniguration for low current measurement applications Floating DUT coniguration Figure 1 shows a circuit block diagram using the loating ammeter with a DC volt...

Страница 4: ...ovides the following beneits The ability to synchronize the voltage source output with the measurement timing The ability to display the voltage and current relationship in an X Y graph on the instrum...

Страница 5: ...at its source or to ilter out the noise from the data Of course if the noise is symmetrically distributed then you can simply use the average of the histogram Figure 4 shows an example of this proced...

Страница 6: ...using the B2981A B2983A Femto Picoammeter requires some modiications to the software key sequences and steps shown in this example although the ammeter setup is identical Of course when using the B29...

Страница 7: ...then follow the steps shown below to set the output voltage to zero volts Step 3 1 Press the Voltage Source On Off key to enable the voltage source to output 0 V Note If using the B2981A 83A then set...

Страница 8: ...nge the voltage to approximately 1 0 V Step 6 1 Continue rotating the Knob and increasing the voltage until the current is between 10 to 20 mA Note If using the B2981A 83A then increase the external v...

Страница 9: ...the B2981A 83A and the voltage applied by the external voltage source on a PC to make this same type of I V plot The following steps outline how to plot the I V characteristics of an LED Step 1 1 Pre...

Страница 10: ...s the Show Trigger assist key The trigger sub panel is now displayed If the Trigger mode is set to MANUAL as shown in step 5 then perform the following steps 5 Using the Knob place the cursor over the...

Страница 11: ...ance and stable picoammeter that can measure down to 100 aA with 10 aA effective resolution The B2980A Series employs a loating ammeter design which provides considerable freedom to use the ammeter in...

Страница 12: ...rs www keysight com find channelpartners Get the best of both worlds Keysight s measurement expertise and product breadth combined with channel partner convenience www keysight com find precisionMEASU...

Отзывы: