background image

05 | Keysight | Low Current Semiconductor Measurements

 

Using the B2980A Series Ammeter - Technical Overview

3. Measurement speed, iltering/averaging and zero cancelation 

functions improve measurement data quality 

(continued)

Null (Zero cancel) capability: 

To remove the effects of any residual offset currents, you can have the instrument 
automatically subtract out these currents by pressing the Null key.

Figure 3 shows a measurement example using the B2980A Series ammeter. In this 
example the B2985A is in its 2 pA measurement range, the measurement speed is set 
to “Quick”, the iltering is set to a 100 point moving average and stray leakage currents 

have been removed using the NULL function. Using the B2985A’s Roll View the data is 
displayed over a 100 second time span. With these settings it is easy to see that a stable 
100 aA level measurement can be made.

4. Histogram View enables real-time data distribution analysis

In order to reduce or eliminate measurement noise it is helpful to be able to analyze 

measurement data statistically in real-time. Once you know the characteristics of the 
noise you are seeing you can decide if it is better to reduce the noise at its source or to 

ilter out the noise from the data. Of course, if the noise is symmetrically distributed then 

you can simply use the average of the histogram. 

Figure 4 shows an example of this procedure using the B2980A’s Histogram View. The 
two histograms show the profound effect that measurement speed and data iltering can 

have when making measurements in a noisy test environment. 

Note: This measurement was made using a banana clip lead to triaxial cable with the input terminal 
open and without shielding; therefore, there is a considerable amount of noise entering the ammeter.

Figure 4(a) shows Histogram View results using 0.1 power line cycle (PLC) measurement 

speed and no iltering. The histogram distribution exhibits peaks at both ends and the 

measured sigma is 0.17 pA.  Figure 4(b) shows the same measurement performed using 

Normal speed (which does PLC averaging) and a 100 point moving average ilter on the 

measurement data. In this case the measured sigma is 0.018 pA, which is about 9.5 times 
(= 0.17/0.018) smaller than the sigma shown in Figure 4(a).

Figure 3. Low current measurement example (B2985A).

The Roll View graph shows the measurement is 
stable at the 100 aA level within a 100 sec interval. 

Measurement condition:

- Speed: Quick
- 2 pA range
- Filter: 100 moving average
- Null: zero cancel
- DUT: Open condition

Quick measurement speed

Zero cancel   Filtering

Содержание B2981A

Страница 1: ...ght Technologies Low Current Semiconductor Measurements Using the B2980A Series Ammeter Keysight B2981A B2983A Femto Picoammeter Keysight B2985A B2987A Electrometer High Resistance Meter Technical Ove...

Страница 2: ...lexibility 3 The ability to improve data quality through averaging iltering and zero cancelation 4 Integrated capability to analyze measurement data via histograms time domain graphs etc The Keysight...

Страница 3: ...rovide stable measurements and a lexible coniguration for low current measurement applications Floating DUT coniguration Figure 1 shows a circuit block diagram using the loating ammeter with a DC volt...

Страница 4: ...ovides the following beneits The ability to synchronize the voltage source output with the measurement timing The ability to display the voltage and current relationship in an X Y graph on the instrum...

Страница 5: ...at its source or to ilter out the noise from the data Of course if the noise is symmetrically distributed then you can simply use the average of the histogram Figure 4 shows an example of this proced...

Страница 6: ...using the B2981A B2983A Femto Picoammeter requires some modiications to the software key sequences and steps shown in this example although the ammeter setup is identical Of course when using the B29...

Страница 7: ...then follow the steps shown below to set the output voltage to zero volts Step 3 1 Press the Voltage Source On Off key to enable the voltage source to output 0 V Note If using the B2981A 83A then set...

Страница 8: ...nge the voltage to approximately 1 0 V Step 6 1 Continue rotating the Knob and increasing the voltage until the current is between 10 to 20 mA Note If using the B2981A 83A then increase the external v...

Страница 9: ...the B2981A 83A and the voltage applied by the external voltage source on a PC to make this same type of I V plot The following steps outline how to plot the I V characteristics of an LED Step 1 1 Pre...

Страница 10: ...s the Show Trigger assist key The trigger sub panel is now displayed If the Trigger mode is set to MANUAL as shown in step 5 then perform the following steps 5 Using the Knob place the cursor over the...

Страница 11: ...ance and stable picoammeter that can measure down to 100 aA with 10 aA effective resolution The B2980A Series employs a loating ammeter design which provides considerable freedom to use the ammeter in...

Страница 12: ...rs www keysight com find channelpartners Get the best of both worlds Keysight s measurement expertise and product breadth combined with channel partner convenience www keysight com find precisionMEASU...

Отзывы: