LAB1: Source Finite Current/Voltage Sinusoid Waveform
Objective
This demo illustrates the arbitrary
waveform generation function by
sourcing a current/voltage sinusoid
waveform and monitoring the outputs
using an oscilloscope.
Procedure
1. Source finite current/voltage
sinusoid waveform
2. (Optional) Configure phase
marker output
3. Monitor the entire output signal
on its GUI
4. View the list of measurement results
Demonstration
1-1. Source a finite current/voltage sinusoid waveform
1) Press
repeatedly until the Channel 1
Single View
is displayed.
2) Press to edit the
Source function
, and then select to set the
Source function
to
Voltage source
. (If you selected to set the
Source function
to
Current source
then it
would source a current waveform.)
3) Press to edit the
Source value
, and enter
0.5 V
to set the
Source value
to
0.5 V.
4) Press to edit the
Limit value
,
and enter
10 mA
to set the
Limit value
to
10 mA
.
5) Press to configure the
Monitor parameter
,
and select to set the
Monitor parameter
to
Current
.
Note:
Although this guide shows a voltage
sinusoid waveform example, you
can easily source a current sinusoid
waveform by selecting Current Source
Mode in step 2.
06 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide