![Keysight Technologies B2960A Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/keysight-technologies/b2960a/b2960a_demo-manual_1962179005.webp)
05 | Keysight | Using an External Trigger to Generate Pulses with the B2960A – Demo Guide
LAB: Using external trigger signals and B2960A series power source to analyze a
device-under-test (DUT)
Objective
This demo illustrates the ability of
a B2960A series power source to
respond to an external trigger with
pre-specified settings: Pulse voltage,
pulse delay and pulse width
Procedure
1. Change View mode to Channel 1
Single View
2. Configure the pulse condition
3. Configure the trigger condition
4. Disable an internal source wait time
5. Initiate the channel to wait for an
external trigger signal
6. (Optional) Abort the channel to stop
waiting for an external trigger signal
Demonstration
1. Change View mode to Channel 1 Single View
1) Press repeatedly until the Channel 1
Single View
appears on the screen.
2. Configure the pulse condition
1) Press to edit the
Source function
,
and then select to set the
Source
function
to
Voltage source
.
2) Press to edit the
Source value
,
and then ente
r
0 V
to set the
Source value
to
0 V.
3) Press to edit the
Limit value
, and then enter
10 mA
to set the
Limit value
to
10 mA
.
4)
Press to change the keys shown in Assist keys, and then press to
show the
Pulse Sub-Panel
.