Keysight Technologies B2960A Скачать руководство пользователя страница 4

04 | Keysight | Using an External Trigger to Generate Pulses with the B2960A – Demo Guide

Setup

Connection

1. Connect the

 black banana cable

 of the U8201A to the B2960A’s 

Ch1 Low Force

 Terminal.

2. Connect the 

red banana cable

 of the U8201A to the B2960A’s 

Ch1 High Force

 Terminal.

3. Clip the 

end of the 1.1 kΩ

 resistor with the 

black alligator clip

 of the U8201A.

4. Clip the 

other end of the 1.1 kΩ

 resistor with the 

red alligator clip

 of the U8201A.

5. Probe across the 

1.1 kΩ resistor

 to monitor the output using the 

oscilloscope’s Ch1

.

6. Connect the 

Pin 9

 of the B2960A’s Digital I/O Port to the 

external trigger signal source

 and 

the oscilloscope’s Ch2 using the BNC cables, the N1294A-031, wiring, etc.

Oscilloscope setup

1. Set the 

Ch1 vertical scale

 to 

500 mV/div

.

2. Set the 

Ch2 vertical scale

 to 

5 V/div

.

3. Set the 

horizontal scale

 to 

500 μs/div.

1.1 k

Ω

Resistor

To oscilloscope’s input Ch 1 

B2960A

External  Trigger Signal

BNC Cable

BNC Cable

N1294A-031

BNC Cable

To oscilloscope’s input Ch 2 

U8201A

Содержание B2960A

Страница 1: ...Keysight Technologies Using an External Trigger to Generate Pulses with the B2960A B2960A 6 5 Digit Low Noise Power Source Demo Guide...

Страница 2: ...ch as power supplies and sources This demonstration guide highlights the excellent pulsed signal output function and the triggering capability via the B2960A series digital I O signal port The methodo...

Страница 3: ...priate trigger condition and initiating the channel the instrument will be ready to receive an external trigger signal and generate pulsed output Figure 3 shows the external trigger signal requirement...

Страница 4: ...stor with the red alligator clip of the U8201A 5 Probe across the 1 1 k resistor to monitor the output using the oscilloscope s Ch1 6 Connect the Pin 9 of the B2960A s Digital I O Port to the external...

Страница 5: ...able an internal source wait time 5 Initiate the channel to wait for an external trigger signal 6 Optional Abort the channel to stop waiting for an external trigger signal Demonstration 1 Change View...

Страница 6: ...6 Rotate to select Channel 1 Pulse Parameters and set them up as below Peak 2 V Delay 300 s Width 200 s 3 Configure the trigger condition 1 Press to show the Trigger Sub Panel 2 Rotate to edit the Tr...

Страница 7: ...e Source trigger source to EXT9 6 If you aren t on the top of the Function menu press repeatedly to return to the top level 7 Press and then press to open the DIO Configuration dialogue 8 Press to edi...

Страница 8: ...e configured 10 Rotate to select the Input output function and press to edit it Then press to set it to Trigger input 11 Rotate to select the Polarity and press to edit it Then press to set it to Posi...

Страница 9: ...re in order to have the measurements made under the exact source condition as specified However it is required to disable it to source pulses at the exact timing as specified by external trigger signa...

Страница 10: ...to make the channel 1 be ready to receive an external trigger signal Now the B2960A series is ready to receive an external trigger signal Send the external trigger signal to the B2960A series to gener...

Страница 11: ...ly available It can source either voltage or current with 6 5 digits of resolution while also monitoring them which is essential for a variety of measurement applications In addition to an arbitrary w...

Страница 12: ...on Keysight Technologies products applications or services please contact your local Keysight office The complete list is available at www keysight com find contactus Americas Canada 877 894 4414 Bra...

Отзывы: