Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
Parameters
MeasTime
MeasTime defines the data extraction timing (see
) which is the timing to
extract the pulsed IV measurement data (gate voltage, drain voltage, and drain
current) from the oscilloscope measurement result (pulse waveform). Initially, this
value is set to 0.5. Changing this value may cause a measurement error. If you
change this value, check the pulse waveform and choose the point where the pulse
voltage is flat and stable.
Figure 5-9
MeasTime
SmplNum
SmplNum defines the sampling number set to the oscilloscope. Set the value in
order to display the pulse waveform on the oscilloscope screen. A small number will
improve measurement throughput.
Sampling rate of the oscilloscope is set to the value automatically calculated from
the specified pulse width and the specified sampling number.
offset = 0
offset = 0
Gate pulse
Drain pulse
Data extraction timing
meas_time
x Pulse width
V/2
V/2
Pulse width
V: Pulse amplitude, > 5.5 mV, < 4.5 V
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......