Keysight B1507A User’s Guide, Edition 1
1- 23
Introduction
Measurement Resources
Measurement Resources
Keysight B1507A is equipped with the following measurement resources.
•
•
•
•
See Keysight EasyEXPERT
User’s Guide Vol. 1
for the functions available for
Keysight EasyEXPERT software. See
Programming Guide
for the functions
available for the GPIB remote mode.
GNDU - Ground Unit
The ground unit (GNDU) is a 0 V constant voltage source, and used for the
reference of measurement ground. Also the GNDU can sink up to ±4.2 A, so it is
effective for using MPSMU or HVSMU. Figure 1-1 is a simplified circuit diagram
of GNDU.
Figure 1-1
Simplified GNDU Circuit Diagram
GNDU
0 V
Force
Common