Performance Verification
This chapter provides performance test verification procedures for the AC6800B Series instruments:
SHOCK HAZARD
These tests should only be performed by qualified personnel. During the performance of
these tests, hazardous voltages may be present at the output of the unit.
Test equipment required
Type
Specifications
Recommended Model
Digital Voltmeter
Resolution: 10 nV @ 1V
Readout: 8½ digits
Accuracy: 20 ppm
Keysight 3458A
AC Current Monitor
0.01
Ω
+/-200 ppm 25 A
Guildline 7350-.01
AC Current Monitor (AC6804B only) 0.01
Ω
+/-200 ppm 50 A
Guildline 7340-50A
DC Current Shunt
0.1
Ω
30 A
Guildline 9230A-30
DC Current Shunt (AC6804B only)
0.05
Ω
50 A
Guildline 9230A-50
Frequency Counter
Accuracy @1 KHZ < 0.001%
Keysight 53210A
Load Resistors
16
Ω
, 5 A, 400 W minimum (AC6801B)
64
Ω
, 2.5 A, 400 W minimum (AC6801B)
8
Ω
, 10 A, 800 W minimum (AC6802B)
32
Ω
, 5 A, 800 W minimum (AC6802B)
4
Ω
, 20 A, 1600 W minimum (AC6803B)
16
Ω
, 10 A, 1600 W minimum (AC6803B)
2
Ω
, 40 A, 3200 W minimum (AC6804B)
8
Ω
, 20 A, 3200 W minimum (AC6804B)
Current Monitoring Resistor
A four-terminal current monitoring resistor (current shunt) listed above is required to eliminate output
current measurement error caused by voltage drops in the load leads and connections.
Keysight AC6800B Series Operating and Service Guide
177
4 Calibration, Verification, and Service
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...