Performance Test Records
Performance Test Record - AC6801B
Test Description
Minimum Spec.
Results
Maximum Spec.
AC Voltage Programming and Readback Accuracy
155 Vrms @ 60Hz
154.4575
______ V
155.5425
Front Panel Measurement
Vrms - 1.125 V
______ V
Vrms + 1.125 V
155 Vrms @ 500Hz
154.4575
______ V
155.5425
Front Panel Measurement
Vrms - 2.085 V
______ V
Vrms + 2.085 V
310 Vrms @ 60Hz
308.915
______ V
311.085
Front Panel Measurement
Vrms - 2.25 V
______ V
Vrms + 2.25 V
310 Vrms @ 500Hz
308.915
______ V
311.085
Front Panel Measurement
Vrms - 4.17 V
______ V
Vrms + 4.17 V
Voltage Load Regulation
CV Load Effect Low Range @ 60 Hz
Vout - 0.15 V
______ V
Vout + 0.15 V
CV Load Effect Low Range @ 500 Hz
Vout - 0.50 V
______ V
Vout + 0.50 V
CV Load Effect High Range @ 60 Hz
Vout - 0.30 V
______ V
Vout + 0.30 V
CV Load Effect High Range @ 500 Hz
Vout - 1.00 V
______ V
Vout + 1.00 V
RMS Current Measurement Accuracy
Low Range Measurement @ 60 Hz
Iout – 45 mA
______ A
Iout + 45 mA
Low Range Measurement @ 500 Hz
Iout – 75 mA
______ A
Iout + 75 mA
High Range Measurement @ 60 Hz
Iout – 22.5 mA
______ A
Iout + 22.5 mA
High Range Measurement @ 500 Hz
Iout – 37.5 mA
______ A
Iout + 37.5 mA
Frequency Accuracy
Program 500 Hz
499.9 Hz
______ Hz
500.1 Hz
DC Voltage Programming and Readback Accuracy
+219 VDC Output
218.343 VDC
______ VDC
219.657
Front Panel Measurement
Vout - 1.445 VDC ______ VDC Vout + 1.445 VDC
+438 VDC Output
436.905 VDC
______ VDC
439.095
Front Panel Measurement
Vout - 2.89 VDC
______ VDC
Vout + 2.89 VDC
DC Current Readback Accuracy
Low Range Measurement
Iout – 40 mA
______ A
Iout + 40 mA
High Range Measurement
Iout – 20 mA
______ A
Iout + 20 mA
Keysight AC6800B Series Operating and Service Guide
185
4 Calibration, Verification, and Service
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...