VOLTage Subsystem
Voltage commands program the output voltage of the instrument.
Soft limits apply only if the corresponding limit state is on, and they only apply to voltage settings executed after the
limit state is set on. Soft limits do not retroactively apply to the existing voltage setting.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
value
>|MINimum|MAXimum[,
<
low_limit
>|MINimum|MAXimum, <
high_limit
>|MINimum|MAXimum]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MINimum|MAXimum]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
value
>|MINimum|MAXimum
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MINimum|MAXimum]
Sets the immediate or triggered AC voltage level, and optionally the soft limits for the immediate voltage. The
triggered level is the value that is transferred to the output when an output step is triggered. Units are in volts (rms).
Parameter
Typical Return
AC mode 135 V range: 0.0 to 137.5 V
AC mode 270 V range: 0.0 to 275.0 V
AC+DC mode: The total voltage must be within the voltage limits, and the AC+DC peak must be between -389 and
+389 V.
+2.01
Sets the output voltage to 20 V: VOLT 20
Sets the triggered voltage to 10 V: VOLT:TRIG 10
– This command takes one or three arguments. You cannot have a low or high limit without the other.
– This command applies to AC and AC+DC.
– Voltage units (MV, V, KV) are supported.
– The
command sets this parameter to 0.
– When VOLT:LIM:STAT is OFF, only individual range checking for each argument is performed. The <
value
> must
be within its own MINimum-MAXimum range considering VOLTage:RANGe and the existing DC component setting
if in ACDC coupling mode. The typical MAXimum in OUTP:COUP:AC mode is 137.5 or 275.0 VAC, but in ACDC mode
with nonzero DC components the value will be reduced. The lower and upper soft limits must be within their
MINimum-MAXimum range as defined by the active VOLT:RANG setting. Both have the same 0 to +137.5 or +275.0
VAC limits regardless of any operational conditions other than VOLT:RANG.
– When VOLT:LIM:STAT is ON, the range checking described above is performed, and an additional check ensures
that the <
value
> does not go beyond the high and low soft limits.
– The allowable range for MINimum and MAXimum varies depending on whether LIM:STAT is ON or OFF. When
LIM:STAT is OFF, the MINimum and MAXimum values are simply calculated from the VOLT:RANG and the DC
component if it is already set in ACDC mode.
Keysight AC6800 Series Operating and Service Guide
151