Keysight 86120D and 86122C Multi-Wavelength Meters User’s Guide
1
Getting Started
Step 5. Connect the Line Cord
Connect the line cord as shown in the figure below:
Figure 3
The product can operate from any single-phase AC power source
supplying 100 – 240 V in the frequency range from 50 – 60 Hz. The
maximum power consumption is 240 W. The power supply automatically
adapts to the applied AC power (Auto Selection) and monitors the AC
power range. The mains supply voltage fluctuation shall not exceed 10% of
the specified supply voltage.
CAUTION
The product must be operated from a power source providing Line,
grounded Neutral and Protective Earth (PE) connections. The instrument
is not intended for operation from two phases of a multiphase ac system
or across the legs of a single-phase, three-wire AC power system. Crest
factor (ratio of peak voltage to rms.) should be typically within the range
of 1.3 to 1.6 at 10% of the nominal rms. mains voltage.