![Keysight Technologies 85309B LO/IF Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/keysight-technologies/85309b-lo-if/85309b-lo-if_users-and-service-manual_1962201046.webp)
3-2 Keysight 85309B User’s and Service Guide
General Information
General Specifications
3-
Nominal Voltage Range
100/120/220/240 Vac (Fluctuation should not /- 10% of
nominal supply voltage)
Power
300 Watts
Weight and Dimensions:
Net Weight
Std: 9.1 kg
Option 001: 9.7 kg
Option 002: 10.3 kg
Option 400: 9.9 kg
Option 401: 10.65 kg
Option 402: 11.35 kg
Dimensions
Height: 146 mm
Width: 426 mm
Depth: 529 mm
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The
ambient temperature (outside the cabinet) must be less than the maximum
operating temperature of the instrument by 4
°
C for every 100 watts
dissipated in the cabinet. If the total power dissipated in the cabinet is
greater than 800 watts, forced convection must be used.
Содержание 85309B LO/IF
Страница 6: ...vi Keysight 85309B User s and Service Guide...
Страница 18: ...1 10 Keysight 85309B User s and Service Guide Getting Started Getting Help with Your 85309B 1...
Страница 44: ...2 26 Keysight 85309B User s and Service Guide Operating Information Performing the Operator s Check 2...
Страница 59: ...Keysight 85309B User s and Service Guide 4 5 Service Replaceable Parts Figure 4 2 Cable Locations Standard...
Страница 63: ...Keysight 85309B User s and Service Guide 4 9 Service Replaceable Parts Figure 4 4 Cable Locations Option 001...
Страница 68: ...4 14 Keysight 85309B User s and Service Guide Service Replaceable Parts 4 Figure 4 6 Cable Locations Option 002...
Страница 73: ...Keysight 85309B User s and Service Guide 4 19 Service Replaceable Parts Figure 4 8 Cable Locations Option 400...
Страница 79: ...Keysight 85309B User s and Service Guide 4 25 Service Replaceable Parts Figure 4 10 Cable Locations Option 401...
Страница 85: ...Keysight 85309B User s and Service Guide 4 31 Service Replaceable Parts Figure 4 12 Cable Locations Option 402...
Страница 86: ...4 32 Keysight 85309B User s and Service Guide Service Replaceable Parts 4...
Страница 87: ...Installation Note Xxxxx xxxxx 3...