Keysight 85059B Operation and Service Guide 3-19
Use, Maintenance, and Care of the Devices
Disconnection Procedure
Disconnection Procedure
To avoid lateral (bending) force on the connector mating plane surfaces,
always support the devices and connections.
1.
Use an open–end wrench to prevent the device body from turning.
2.
Use another wrench to loosen the connector nut.
3.
Complete the disconnection by hand, turning only the connector nut.
4.
Pull the connectors straight apart without twisting or bending.
Handling and Storage
— Install the protective end caps and store the calibration devices in the
foam-lined storage case when not in use.
— Never store connectors loose in a box, desk, or bench drawer. This is the
most common cause of connector damage during storage.
— Keep connectors clean.
— Do not touch mating plane surfaces. Natural skin oils and microscopic
particles of dirt are easily transferred to a connector interface and are very
difficult to remove.
— Do not set connectors contact-end down on a hard surface. The plating and
the mating plane surfaces can be damaged if the interface comes in
contact with any hard surface.
Do
not
twist one connector out of the other, (like removing a light bulb).
Turn the connector nut, not the device body. Major damage to the center
conductor and the outer conductor can occur if the device body is twisted.
Содержание 85059B
Страница 1: ...Keysight Technologies 85059B 1 0 mm Precision Calibration Kit Operating and Service Manual...
Страница 6: ...4 Keysight 85059V Operation and Service Guide Contents...
Страница 22: ...1 16 Keysight 85059B Operation and Service Guide General Information Regulatory and Environmental Information 1...
Страница 48: ...3 20 Keysight 85059B Operation and Service Guide Use Maintenance and Care of the Devices Handling and Storage 3...
Страница 66: ...85059V Index Index 4 part numbers 6 2 6 4...
Страница 67: ...Installation Note Xxxxx xxxxx 3...